Ieškoti

Vittamed 205 Non-invasive Intracranial Pressure (ICP) Meter with CE Mark
Aprašymas:

Vittamed 205 applies controlled external pressure (Pe) to the tissues surrounding the eye ball and uses two-depth Doppler ultrasound to compare simultaneous blood flow parameters in two contiguous segments of the ophthalmic artery (OA) – the intracranial and extracranial segments.

The device directly compares ICP and Pe values and can identify the balance ICP=Pe when blood flow pulsatilities in both OA segments are similar.

Novelty:

First non-invasive intracranial pressure meter capable of accurately, precisely and reliably measure the absolute value of intracranial pressure (ICP) without the necessity of periodic calibration to the individual patient.

Raktažodžiai: Vittamed 205, Non-invasive Intracranial Pressure (ICP) Meter

Mokslo kryptis: Diagnostinės ir matavimo technologijos, Medicina